Abstract:Abstract: Objective To improve the clinical safety and clinical effectiveness of JZ-1C CO2 laser treatment instrument by measuring its parameters including the power of indicator light, time stability, reproducibility and relative deviation of indicating value (δp). Methods Measured the above parameters by using AD7888 analog-digital converter. Results The measured values of the above parameters can meet the requirements of national mandatory standards in China so that the JZ-1C CO2 laser treatment instrument can be used in clinical sequentially. Conclusion The effective quality control of CO2 laser treatment instrument should be executed to avoid the clinical risks and improve the clinical effectiveness.
Key words: laser treatment instrument; quality control; the power of indicator light; time stability; reproducibility; relative deviation of indicating value