Abstract:In order to research the influences of application of artificial hand and test set-up to EMC tests, studies were performed on
the EMC product Class Standard YY 0505-2012, the General Standard GB 4824-2013 and the basic standard GB/T 6113.102-2008.
Relevant test verification was conducted on the electrophysiological diagnostic equipment, which including conducted emissions,
conducted RF disturbance and electrical fast transient/burst immunity, we found that the application of the artificial hand and the test
set-up could not be ignored. It is helpful for testing specification of EMC test to understand the principle of artificial hand correctly
by analyzing the reason of influence of artificial hand.
刘美丽,张淑溢,王晓慧,李庆雨,施燕平. 电生理诊断中模拟手的应用[J]. 中国医疗设备, 2020, 35(3): 14-16.
LIU Meili, ZHANG Shuyi, WANG Xiaohui, LI Qingyu, SHI Yanping. Application of Artificial Hand in Electrophysiological Diagnostic Equipment. China Medical Devices, 2020, 35(3): 14-16.