Abstract:Objective To investigate the accuracy of non-invasive blood pressure monitor in blood pressure measurement of patients with atrial fibrillation. Methods The blood pressure of 60 patients with atrial fibrillation was measured by using non-invasive blood pressure monitor and mercury sphygmomanometer, respectively. Then the measuring results of blood pressure of non-invasive blood pressure monitor were compared with those of mercury sphygmomanometer. Results There was no significant difference between the measuring results of blood pressure of non-invasive blood pressure and those of mercury sphygmomanometer on systolic pressure (P>0.05) while there were significant differences between the measuring results of blood pressure of non-invasive blood pressure and those of mercury sphygmomanometer on diastolic pressure (P<0.05). Conclusion It is inaccurate to measure the blood pressure of patients with atrial fibrillation by using non-invasive blood pressure monitor, which should be paid attention to in clinic.
胡文志,周海波. 无创血压监护仪在心房纤颤病人血压测量中的准确性评价[J]. 中国医疗设备, 2014, 29(4): 120-122.
HU Wen-zhi, ZHOU Hai-bo. Accuracy Evaluation of Non-invasive Blood Pressure Monitor in Blood Pressure Measurement of Patients with Atrial Fibrillation. China Medical Devices, 2014, 29(4): 120-122.