Influence of Calibration on Testing Results of Mixed and Dead Volume Reagents of Electrochemical Luminescence
LI Guang-quan1, ZHOU Wei-dong1, MA Yi-yun2, ZHANG Lin1
1.Department of Clinical Laboratory, The 363th Hospital of Chengdu, Chengdu Sichuan 610041, China; 2.Rongsheng Pharmaceutical Co.,Ltd., Chengdu Sichuan 610041, China
Abstract:Objective To explore the influence of calibration on testing results of mixed and dead volume reagents of Roche electrochemical luminescence. Methods To test PCT values of quality control materials and patients (20 cases) with mixed and dead volume reagents before and after calibration. Then the testing results before and after calibration were compared according to the testing results of a box of new reagent. Results There were significant differences between the testing results before and after calibration (P<0.05), which showed that the testing precision of mixed and dead volume reagents after calibration was higher than that of mixed and dead volume reagents before calibration. Conclusion The mixed and dead volume reagents of Roche electrochemical luminescence should be calibrated and then can be used in clinic.
李广权,周卫东,马异云,张林. 校准对电化学发光混合死腔量试剂检测结果的影响[J]. 中国医疗设备, 2014, 29(4): 116-117.
LI Guang-quan, ZHOU Wei-dong, MA Yi-yun, ZHANG Lin. Influence of Calibration on Testing Results of Mixed and Dead Volume Reagents of Electrochemical Luminescence. China Medical Devices, 2014, 29(4): 116-117.