Abstract:Objective To measure and analyze the occlusion alarm parameters of infusion pump under different infusion conditions to
find out the quality control rules of the occlusion alarm of infusion pump. Methods IDA-4 Plus infusion pump tester (Fluke, USA)
was used to test and analyze the occlusion alarm parameters of infusion pumps under different infusion pipelines, different infusion
liquids, different flow rates, and different setting thresholds of infusion pump. Results Through the experimental analysis, the
results showed that the flow rate of the infusion pump, the selection of the infusion pipelines, different setting pressure thresholds of
infusion pump, and the infusion liquid with different viscosity all had effects on the parameters of the infusion pump occlusion alarm.
Conclusion This paper analyzes the influencing factors of infusion pump occlusion alarm parameters, which has important guiding
significance for the safe use of infusion pump in clinic.
北京积水潭医院 医疗设备科,北京 100035. 输液泵阻塞报警的质控研究[J]. 中国医疗设备, 2019, 34(11): 39-43.
HAN Qian. Quality Control Research of Infusion Pump Occlusion Alarm. China Medical Devices, 2019, 34(11): 39-43.