Abstract:In this paper, the failure of no X-ray and immovable of C-arm of GE Senographe DS mammography machine was
analyzed, so as to provide reference for medical engineers to repair similar failure. According to two kinds of failure phenomenon,
the faults were found and solved by reviewing the system event log and maintenance manual. After a week of running, all indexes
were normal. The respective method was feasible and effective for both malfunction above, and could reduce the maintenance cost.
熊贵富. GE Senographe DS乳腺机维修案例分析[J]. 中国医疗设备, 2019, 34(4): 171-173.
XIONG Guifu. Maintenance Case Analysis of GE Senographe DS Mammography Machine. China Medical Devices, 2019, 34(4): 171-173.