Abstract:Due to the working principle and characteristics of infusion pump, flow detection, especially the flow detection which is out of the scope of technology standard would take longer to be finished in the detection process of infusion pump quality control. Based on the original detection accuracy, a series of experiments were conducted during the detection process of infusion pump quality control. The detection methods of unqualified infusion pump were adjusted by analyzing the experimental data in practical work. Thus the amount of ineffective work time have been reduced and the detection efficiency of infusion pump have been improved.
许耀良,于洋,陈基明. 输液泵质量检测技术规范中测试时间选择问题的探讨[J]. 中国医疗设备, 2014, 29(5): 55-57.
XU Yao-liang, YU Yang, CHEN Ji-ming. Discussion on Problems of Testing Time Selection in Technology Standard for Quality Control of Infusion Pump. China Medical Devices, 2014, 29(5): 55-57.