Abstract:This paper analyzed the original installation structures of the encoder malfunctioned in the scanning process of CT bed, and pointed out that the unparallel installation of various adapting pieces of the encoder was the failure cause. Then this paper introduced the floating installation process of the encoder as well as corresponding performance test results. The failure of the encoder was solved with the application of the floating installation process.
王迪. CT床编码器浮动式安装与测试[J]. 中国医疗设备, 2014, 29(7): 128-130.
WANG Di. Floating Installation and Performance Testing of CT Bed Encoder. China Medical Devices, 2014, 29(7): 128-130.