Abstract:According to the characteristics of central air supply and air supply from the pump of the air cushion bed, the YQ-P series medical jet air cushion bed was improved through addition of an air-pressure-reduced valve between central air supply and the air bed mattress so as to make the output pressure adjustable from 0 to 0.03 MPa. Adoption of central air supply instead of the air pump in the hospital eliminated various problems that might be caused by the air pump, such as faults, noise and pressure instability. This improved YQ-P series medical jet air cushion bed demonstrated its convenient safe, easy-to-operate and space-efficient features in more effective prevention and treatment of pressure sores for critically-ill patients and realization of resource sharing in the hospital.
王明亮,郑志满,胡昕,伍锡勇,张冕. YQ-P系列医疗用喷气气垫床供气装置的改进[J]. 中国医疗设备, 2015, 30(5): 112-113.
WANG Ming-liang, ZHENG Zhi-man, HU Xin, WU Xi-yong, ZHANG Mian. Improvement of the YQ-P Series Medical Jet Air Cushion Bed. China Medical Devices, 2015, 30(5): 112-113.