Clinical Application of Driven Equilibrium Technology in MRI T2-Weighted Image of Ankle
LIU Shu-hua1, LIU Xue-ming1, PANG Pei-pei2, GU Chang-chun2
o MRI T2 weighted image (T2WI) can obviously reduce scanning time and improve scanning speed on the premise of ensuring preferable image contrast and image quality.
摘要目的 通过对磁共振常规T2WI施加驱动平衡技术,探究该技术在保证图像对比度和图像质量的前提下能否显著缩短扫描时间、加快扫描速度。方法 随机选取10名志愿者对其踝关节进行磁共振检查。每名志愿者进行3种T2WI扫描,分别为常规T2WI扫描(TR=4250 ms),短TR时T2WI扫描(TR=2500 ms),短TR时且施加驱动平衡的T2WI扫描(TR=2500 ms)。分析各种扫描图像的软组织对比度和图像质量。结果的部分未复制上。结论 在T2WI中施加驱动平衡技术,可在保证图像质量的前提下缩短扫描时间,值得推广。[关键字] 踝关节;驱动平衡技术;T2WI;图像对比度Abstract: Objective To test whether applying the driv
Abstract:Objective To test whether applying the driven equilibrium technology to MRI T2 weighted
image (T2WI) can obviously reduce scanning time and improve scanning speed on the premise of ensuring
preferable image contrast and image quality. Methods Ten volunteers were selected randomly to undergo
MRI examinations on their knees. All the volunteers received three kinds of T2WI scan: conventional T2WI
scan (TR=4250 ms), T2WI scan with short TR (TR=2500 ms), T2WI scan with driven equilibrium technology
and short TR (TR=2500 ms). Then the soft-tissue contrasts and quality of images were contrastively analyzed.
Results With the application of driven equilibrium technology, the contrast and quality of images could be
improved and the scanning time could be significantly reduced. Conclusion Applying driven equilibrium
technology to T2WI can reduce scanning time on the premise of ensuring image quality, which is worth widely
promotion.
刘树华,刘学明,庞佩佩,谷长春. 驱动平衡技术在踝关节磁共振T2WI成像中的临床应用[J]. 中国医疗设备, 2015, 30(1): 60-62.
LIU Shu-hua, LIU Xue-ming, PANG Pei-pei, GU Chang-chun. Clinical Application of Driven Equilibrium Technology in MRI T2-Weighted Image of Ankle. China Medical Devices, 2015, 30(1): 60-62.
Gold GE,Fuller SE,Hargreaves BA,et al.Driven equilibrium magnetic resonance imaing of articular cartilage:initial clinical experience[J].J Magn Reson Imaging,2005,21(4):476-481.
[6]
Bellon EM,Haacke EM,Coleman PE,et al.MRI artifacts:a review[J].AJR Am J Roentgenol,1986,147:127-1381.
[2]
王逗.核磁共振原理及其应用[J].现代物理知识,2005,17(5):50-51.
[8]
Thomas DL,Pell GS,Lythgoe MF,et al.A quantitative method for fast diffusion imaging using magnetization-prepared TurboFLASH[J].Magn Reson Med,1998,39(6):950-960.