Abstract:Objective To explore the clinical application value of a new-type and disposable anti-bacteria central venous blood collector. Methods 50 patients were involved in this study. Central venous blood collection of the patients was conducted with disposable anti-bacteria central venous blood collector (observation group) while peripheral venous blood collection of the patients was conducted with disposable vacuum venous blood collector (control group). The one-time success rate of blood collection, inspection results and complications of observation group were compared with those of control group. Results The one-time success rate of blood collection of observation group was 98%, which was significantly better than that of control group (84%) (P<0.05). There was no significant difference between observation group and control group in terms of inspection results (P>0.05). Blood culture results of observation group showed that there was no bacteria exist in blood samples. The complication incidence rate of observation group was 0, which was significantly better than that of control group (10%) (P<0.05). Conclusion Disposable anti-bacteria central venous blood collector which owns advantages such as safety and efficacy should be promoted in clinic.
王清凤,谷欣. 新型一次性防菌型中心静脉采血器的临床应用[J]. 中国医疗设备, 2014, 29(11): 126-128.
WANG Qing-feng, GU Xin. Clinical Application of New-Type and Disposable Anti-Bacteria Central Venous Blood Collector. China Medical Devices, 2014, 29(11): 126-128.