Abstract:Objective To evaluate the quality of dialyzer of different brands via studying the quality control method and cycle of
blood dialyzer. Methods A total of 30 installed hemodialysis machines of two brands were selected from the hemodialysis center
of Yongchuan Hospital, Chongqing Medical University. The conductivity, temperature, and other performance indicators of the
hemodialysis machines were tested at the end of each quarter, and the results were analyzed. Results The errors in conductivity
and temperature had significant differences at various time points (P<0.05). The error of venous pressure was significantly different
between the first quarter and the third and fourth quarters (P<0.05), and there was a significant difference between the second quarter
and the third and fourth quarters (P<0.05). The errors in arterial pressure and dialysate flow rate were significantly different in the
first quarter from the second, third, and fourth quarters (P<0.05), and there was a significant difference between the second quarter
and the third and fourth quarters (P<0.05). In the third and fourth quarter of the domestic models and imported models hemodialysis
machine, there was a significant difference in conductivity and temperature (P<0.05). Conclusion The deviation of each performance
index is significantly increased in one year’s detection. Imported models hemodialysis machine is superior to domestic models in
stability, but the performance indexes of the two models can be similar by testing and calibrating.
刘晏明,熊贵富,廖娟. 血液透析机质量控制的研究及探讨[J]. 中国医疗设备, 2018, 33(9): 49-52.
LIU Yanming, XIONG Guifu, LIAO Juan. Research and Discussion on Quality Control of Hemodialysis Machine. China Medical Devices, 2018, 33(9): 49-52.