Abstract:The stability and reliability of ventilator is of great significance. However, the standard of quality control does not claim
any requirement for the control precision of ventilator. The present study designed an experiment to test the control precision of
pressure control mode when the load of ventilator changed. The results showed that when the respiration parameters changed, the
control precision of ventilator was influenced. It sometimes failed to meet corporate standard, even decreased the safety of ventilator,
which might cause lung injury.
侯晓旭,李佳戈,任海萍. 气道阻力及肺的顺应性对呼吸机压力控制精度的影响[J]. 中国医疗设备, 2018, 33(8): 38-41.
HOU Xiaoxu, LI Jiage, REN Haiping. Effect of Airway Resistance and Lung Compliance on the Pressure Control Precision of Ventilator. China Medical Devices, 2018, 33(8): 38-41.