Abstract:Ventilator is a kind of emergency equipment which is necessary for first aid in hospital, and it is commonly used in clinical
adjuvant. This machine can strive for valuable treatment time for extending patient’s life, and it can create good treatment conditions
for patient’s treatment. Siemens MAQUET Servo-s ventilator is designed for tracheotomy and noninvasive ventilation support (via
face mask) of adult patients. In the present study, the pre-use self-inspection function and fault self-checking function of MAQUET
Servo-s were introduced, and the cause of the fault by fault code “technique fault 5” and failure phenomenon was found. The circuit
board was repaired by use of professional equipment and professional knowledge, the maintenance ability of engineer was enhanced
and the maintenance fund was saved.
贾立民,张鹏飞. MAQUET Servo-s呼吸机的故障分析及维修一例[J]. 中国医疗设备, 2018, 33(3): 101-103.
JIA Limin, ZHANG Pengfei. One Cases of Troubleshooting for MAQUET Servo-s Ventilator. China Medical Devices, 2018, 33(3): 101-103.