Case Analysis of GE LOGIQ E9 Ultrasonic Diagnostic Equipment Maintenance
DI Haiyan1,LI Ya2,BAI Fangfang1,LI Ya1
1. Department of Functional Section, Baoding No.2 Central Hospital, Baoding Hebei 072750, China丨 2. Department of Ultrasound,
Mancheng People’s Hospital, Baoding Hebei 072150, China
Abstract:This article introduced the structure and principle of the ultrasound equipment GE LOGIQ E9, and explained three
examples of faults, including slow booting of the device, no echo in the image scanning area of the device and restarting of the
device, which fault one was the front-end part fault, the second fault was the back-end part fault, and the third fault was starting up
and restarting failure. The fault was locked in the front or back area through the structural principle of the device, and the cause of the
fault is found row by row through the front and rear boards. Finally, the fault was eliminated by replacing the board, power supply,
remove the back end processor and BIOS troubleshooting. By summarizing the common maintenance ideas and methods of related
cases of the machine during and after the start-up, it can help engineers quickly solve problems according to the fault phenomenon
and ensure the normal operation of the equipment.
邸海燕1,李娅2,白芳芳1,李娅1. GE LOGIQ E9超声诊断设备维修案例分析[J]. 中国医疗设备, 2021, 36(7): 178-180.
DI Haiyan1,LI Ya2,BAI Fangfang1,LI Ya1. Case Analysis of GE LOGIQ E9 Ultrasonic Diagnostic Equipment Maintenance. China Medical Devices, 2021, 36(7): 178-180.